ind_ti17c_250.jpg 
 
 
 
 
 
 
 
  
quikcal_zero_lr.jpg
 
 
 
quikcal_cal_lr.jpg

 

What is Quik-Cal ?

Quik-CalTM The Original Push-Button Zero and Calibration Set since 2007

One of the labor-saving features of DFE's tension amplifier, indicator and controller products is Quik-CalTM technology. It's a feature integrated into almost all of DFE's electronic products:

ind_iamp2-grp_150.jpg           sw5_main.jpg         fg2_openxencl_250.jpg

 

 

With Quik-CalTM the indication and control electronics use front-panel push buttons instead of potentiometers for zero and calibration settings. See the example on the TI17C amplifier to the left.

- No tension display is needed.

- No screwdriver is needed.

- No second person is needed for setup or calibration.

Each button is pressed once during the calibration procedure for one second and the settings are stored automatically.

Most of DFE's tension indication and control products use Quik-CalTM including the TrueTension TI17C and TI18C amplifiers. The TI17C is shown here to illustrate the ease-of-use in completing the ZERO and CALIBRATION procedure once the indicator is mounted and connected to tension transducer(s).

ZERO Set

Because an output signal is generated by the weight of the un-webbed transducer roll in the system, we can negate the roll weight output component in a procedure called ‘zeroing'.  To set the indicator output to ZERO when there is no tension, the ZERO button is pressed for one second.

With Quik-CalTM the zeroing step can even be eliminated altogether, if desired. The tension indicator can be adjusted so that it automatically sets the output to ZERO when power is turned on. This is achieved by changing a jumper setting on the circuit board.

 

CAL Set

Once the indicator is zeroed, calibration is completed by hanging a known weight on a cord between the idler rolls in the machine web path such that the wrap angle over the transducer roll is kept identical to the one used in operation.

When the CAL button on the front of the indicator is pushed and held for one second, the calibration setting is stored. The indicator automatically multiplies the calibration weight by the preset calibration ratio to calculate the full-scale output of the indicator, and the indicator is calibrated and ready for operation.

 

What's Calibration Ratio?

The calibration ratio is a multiplier value that the indicator uses to calculate the tension required to generate the indicator's full output. It can be one of two preset values. The user can choose either 10%, which is the default value, or 25%, which may be preferred for lower tension measurement. The calibration weights chosen by the user will represent either 10% or 25% of the maximum tension the machine is expected to be running.     

The standard calibration ratio is 10% (1:10).  So, if a 15-pound weight is hung for calibration, and the CAL button is pushed, the indicator will store that value as 10% of its full output. Then 150 pounds of tension on the transducer roll (15 x 10) during operation will result in the indicator producing its full-scale output.

If the calibration ratio of 25% is selected instead, the same 15-pound calibration weight will represent 25% of the indicator's full output. So, then, only 60 pounds of tension (15 x 4) on the transducer roll during operation would generate the indicator's full output signal.

To switch between the 10% and 25% calibration ratios, a jumper is moved on the circuit board.

 

Stability

Another benefit of Quik-CalTM technology is inherent stability. The ZERO and CALIBRATION settings are stored digitally, so there is no drift over time and temperature variations as there can be with potentiometers.

 

 

 

 

 

 

 TOP •  Home •  Products •  News •  Support •  Resources •  Customers •  About DFE •  Contact Us